Single-phase transformers for medical use

download PDF file
Type:
associated transformer not resistant to short circuit
Primary voltage: 0/230V
Secondary voltage: 0/230V (115/0/115V)
Transformers for incorporating and installing in electrical plants for medical use (CEI 64-4; CEI 62-5)
Double copper screening inserted between the primary and secondary windings, connected to a metallic support or on a clamp
Reference standards: CEI 96-16; EN 61558-2-15

PRODUCT NOT INTERESTED BY 204/108/CE NORM RELATED TO ELECTROMAGNETIC COMPATIBILITY EMISSIONS.

PART
NUMBER

POWER
[VA]

Height [mm]

Width [mm]

Depth [mm]

Weight [kg]

Mounting  A [mm]

Mounting
B [mm]

Drawing

BTTMIM033.1

500

155

155

120

10

132

83

1

BTTMIM048.1

1000

185

185

160

18

130

128

1

BTTMIM057.1

1500

200

200

180

21

175

138

1

BTTMIM063.1

2000

200

200

210

26

175

168

1

BTTMIM072.1

3000

300

240

220

37

203

135

2

BTTMIM084.1

5000

300

240

240

46

203

155

2

BTTMIM093.1

8000

350

280

250

75

234

193

2

Notes:
1.The total power of the transformer is supplied in relation to the highest secondary voltage.
2.Drawings, measurements and weights are indicative and can change without notice.
3.For powers and voltages not reported in the table or for special construction requirements, request the characteristics of the transformers from our technical office.